I. Principle of UV DOAS
Pollutants such as SO?, NO and NO? have obvious, narrow characteristic absorption structures in the 200–430 nm UV band. DOAS splits the absorption spectrum into two parts:
- Slow-varying part:Caused by Rayleigh scattering, Mie scattering, dust attenuation, varying slowly with wavelength
- Fast-varying part:Caused by target gas characteristic absorption, oscillating rapidly with wavelength
By mathematically high-pass filtering the spectrum, DOAS "peels" the target gas's narrow-band absorption from the broad-band background interference, then fits it against standard cross-sections by least squares to retrieve concentration. This differential idea makes the measurement almost immune to water vapor, dust and other broad-band interference.
II. Why ultra-low emission needs DOAS
After retrofit, coal-fired plant SO? is below 35 mg/m3 and NO? below 50 mg/m3, only about 12–30 ppm by volume. Traditional NDIR or electrochemical methods at this range have strong cross-interference, large drift and slow response.
- High sensitivity:Sub-ppm detection limit, meeting ultra-low-emission threshold judgment
- Multi-component:One analyzer measures SO?, NO, NO? (→NO?), O? and more
- Strong anti-interference:Differential algorithm auto-removes broad-band absorption of water and dust
- No consumables, low maintenance:Long-life UV source, no reagents, suited to long online operation
III. DOAS with long optical path cell
The YTG-2000 series combines DOAS with a long-path multi-reflection cell: within a limited volume, the path is extended to hundreds of mm or meters, increasing the absorption signal and thus the low-concentration SNR. With constant-temperature and constant-pressure control, long-term zero and span stability are ensured.
IV. Typical applications
- Coal-fired / CHP plants:SO?/NO? online monitoring for ultra-low-emission acceptance and routine regulation
- Industrial boilers & kilns:NO?/SO? monitoring for sintering, cement and glass kilns
- Desulfurization/denitrification control:Inlet/outlet comparison to optimize environmental facility operation
- VOC pretreatment & catalytic monitoring:Extend measured components via UV absorption features
V. Data quality and standards
Per HJ75/HJ76 "Technical specifications for flue gas CEMS (SO?, NO?, particulates)", UV DOAS analyzers need periodic zero/span checks, full-range calibration and comparative monitoring. The YTG-2000 series supports auto-calibration, self-diagnosis and remote maintenance, helping enterprises meet environmental requirements while lowering O&M cost.