YTG-2000 Online UV Flue Gas Analyzer
Adopts UV differential absorption spectroscopy (DOAS) and a >2 m long optical path absorption cell for ultra-low-range gas analysis, closely matching current ultra-low-emission environmental requirements.
Product overview

The YTG-2000 series online UV flue gas analyzer uses UV differential absorption spectroscopy (DOAS) and a >2 m long optical path absorption cell to analyze gases at ultra-low ranges, ideal for current ultra-low-emission requirements.
This product selects deep-UV absorption spectra, so the measurement is theoretically unaffected by water interference, which benefits ultra-low-range or ultra-low-emission measurement.
With no moving optical parts — no chopper wheel or interferometer — it offers excellent long-term stability and reliability.
The analyzer's single-board technology uses a high-performance chip for high-speed analysis of sensor data, and its modular design helps reduce product cost.
Features
- Ultra-low-range measurement:UV DOAS + long optical path for high-precision measurement of ultra-low-range components, meeting strict environmental monitoring requirements.
- Water interference immunity:Deep-UV absorption spectra make readings theoretically unaffected by water, ideal for high-humidity environments.
- High reliability:No moving optical parts (no chopper wheel or interferometer), ensuring good long-term stability and reliability.
- Fast response:High-performance single-board chip for high-speed analysis and ultra-low-latency real-time data.
- Modular design:Modular structure for easy maintenance and upgrade, lowering total cost of ownership.
- Multiple outputs:RS485, RS232 and 4-20mA outputs for easy integration; can also analyze NH?, Cl?, H?S and more.
Typical applications
- Ultra-low emission:Continuous SO?/NO? monitoring (CEMS) for coal-fired power plants and industrial boilers.
- High-humidity conditions:Precise measurement in high-humidity flue gas after desulfurization and denitrification.
- Industrial process:Online flue gas composition analysis in chemical, metallurgical and other industries.
Specifications
| Parameter | Gas | Spec |
|---|---|---|
| Range | SO? | 0-50-100-500mg/m3 |
| NO? | 0-50-100-500mg/m3 | |
| O? | 0~25% | |
| Cross-interference | NOx | < ±1%F.S |
| SO? | < ±1%F.S | |
| H?O on NOx | <±1%F.S | |
| H?O on SO? | <0.5%F.S | |
| Detection limit | ≤1% F.S. | |
| Linearity | ±1%F.S. | |
| Repeatability | 1%. | |
| Zero drift | ±0.5%F.S.(1 hour) | |
| Span drift | ±1%F.S.(1 hour) | |
| Operating temp. | -10~40℃ | |
| Sample flow | 0.8L/min | |
| Power | 220VAC, 250W(Max) | |
| Warm-up | ≤ 5 minutes | |
| Flow | 0.8L/min±0.2L/min | |
| Size | 19"*4U*500mm | |
| Weight | 17Kg | |
| Output | RS485, RS232, 4-20mA | |